核心提示:class ii type a1 bsc characteristics exhaust approximately 30% of total air handled and recirculate 70% in cabinet minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening may exhaust hepa filtered air back into the room or through

 

 

class ii type a1 bsc characteristics
exhaust approximately 30% of total air handled and recirculate 70% in cabinet

minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening

may exhaust hepa filtered air back into the room or through a canopy (to the outdoors)

audiovisual alarm required when connected to a building exhaust system

may have contaminated ducts and plenums under positive pressure to the room
type a1 cabinets are not suitable for work with volatile toxic chemical and volatile radionuclides.

 